Contact Us
Hydrogen Production Industry Solutions
It is applied in the synthesis of methane-rich gas with hydrocarbon (Light oil, naphtha, LPG, light hydrocarbon, hydrocarbon-containing dry gas etc.) and steam as raw materials. The resulted gas can be further used to produce hydrogen, syngas with different CO/H2 ratio (for the synthesis of methanol or ammonia etc.) and coal gas.
Product Details
Product Application
NFY-IIA/HFY-3A is an online analyzer for hydrogen content detection in industrial production processes such as water electrolysis for hydrogen and oxygen generation, fertilizer manufacturing, petrochemicals, air separation, and biological fermentation, suitable for both explosion-proof and non-explosion-proof environments.
fundamental
The instrument uses a three-electrode electrochemical sensor, which consists of two circuits: one circuit is the reaction process of the working electrode, the other circuit is the current of the electron transmission. The magnitude of the current of the electron circuit represents the content of
hydrogen in the gas.
Product Features
●The electrochemical sensor has the characteristics of long life, high precision and fast response.
●Small in size and light in weight, easy to install;
●split explosion-proof design;
●Display live;
●The output signal can be transmitted up to 1000 meters.
intrinsically safe
|
Dimensions:
|
140×140×220mm (length×width×height) |
| Cutout dimensions | 136x136(length*width) |
| External safety parameters: | Vo≤28 VDC,lo≤93mA |
| Cp≤0.02uF,Lp≤1mH | |
| Explosion-proof mark | Exiadb IICT4 |
Technical Specifications
| Measuring range | 0.00~2.00%/4.00%(range optional) |
| Maximum allow able error: | ±1%FS |
| Repetitivenes | ±1%FS |
| Resolution | 0.01% |
| Response time | T9o≤10s |
| Zero drift | ≤±%FS/7d |
| Range drift | ≤±1%FS/7d |
| Air flow rate | (300±30)mL/min |
| Working power supply | 24VDC/220VAC |
| work environment | Operating temperature:-20℃ to +45℃ |
| Operating humidity:≤90%R |
Extrinsically safe
| Analog output | 4-20mA |
| numeric output | RS458 interface with Modbus protocol. |
| Explosionproof sign | Exdb II CT6 Gb |
Operation conditions
Catalyst Characteristics
Category: